Standard Parameters
FHR
TOCO
FM
ECG
SpO2
NIBP
Respiration
Temperature
Pulse
Feature Set
12.1 Large-screen Color TFT Display
Light Weight and Portable, Easy Operation Mode
Build-in High-capacity Battery, Up to 5 Hours Working Time
Maximum 11 Hours Storage and Replay of Fetal/Maternal Monitoring Data
Optional Built-in Recorder, Able to Print or Store Waveforms and Monitoring Status at Real Time
Advanced Digital Signal Processing (DSP) Technique is Applied for Real-time Calculation and Mode Identification
High-sensitive Probe, Excellent Signal/Noise Ratio and Accurate Calculation Enable to Monitor Fetus More Safely
Individual Digital Module Structure, Convenient for Maintenance and Upgrade
Digital SpO2 Technique, Precise Measurement for Low Perfusion and Finger Shaking
Protection Against Interference from Defibrillation and High-frequency Electrotome
Intellectualized Voice/Light Alarms Accurately Transmit Diseases
Adult/Pediatric/Neonatal measurement modes
Network Capability, Compatible with Obstetric Central Monitoring System
Standard Accessories
1 x ECG Cable & Wires, 1 x Adult Digital SpO2 Probe, 1 x Adult NIBP Cuff, 1 x TEMP Probe, 1 x Pack Electrodes,
1 x Built-in Battery, 1 x Three in One Probe (FHR, TOCO, FM), 1 x Biding Belt, 1 x Transmission Gel
Optional Accessories
Twin, Built-in Recorder, Special Accessories for Neonatal and Pediatric Patient
Technical Specifications
TOCO
Range: 0-100 pressure units
Nonlinear Tolerance for Measurement: ±5%
FHR
Probe: 9 chips, wide wave packet, Pulse wave
Strength of ultrasonic wave: less than 3mW/cm2
Working frequency: 1.0MHz or 2.2MHz
Signal processing: special digital signal processing (DSP) and mode identification
Range: 10-240bpm
FM
Counting Method: Manual
ECG
Number of Leads: 5
Lead Selection: I, II, III, avR, avL, avF, and V1~V6
Gain Selection: x0.5, x1, x2, and x4
Sweep speed: 12.5mm/s, 25mm/s, and 50mm/s
HR display: 20~250 bpm
Precision: ±1bpm
NIBP
Measurement Technique: Automatic Oscillometric
Measurement Range:
Adult: 20-245mmHg
Pediatric: 20-135 0mmHg
Accuracy: ±5 mmHg
Resolution: 1 mmHg
Measure Mode: Manual /Automatic
Automatic Measuring Interval: 3-240min
Temperature
Range: 20°C-45°C
Accuracy: ±0.1°C
Resolution: 0.1°C
Pulse
Range: 20-250bpm
Accuracy: ±1bpm
Resolution: 1bpm
SpO2
Range: 70-100%
Resolution: 1%
Accuracy: ±1%
Respiration
Range: 0-60bpm
Accuracy: ±1bpm
Resolution: 1bpm
Alarm
Range: Adjustable from Lower to Upper Limit
Cause: Irregular Indication
Indication: Audible and Visual Dual Indication
Safety
In Compliance with IEC60601 Related Standards
Power Requirement
Voltage: AC100-240V 50HZ/60HZ
Printer (Optional)
Type: Built-in 50mm Thermal Printer
Port: Standard Parallel
Weight
6.3kg
Dimension
320mmΧ280mmΧ170mm
|