AV-9100 Single Channel ECG

AV-9100

Model:AV-9100

 

 

 

 

 

 

 

 

 

 

 

 


 

Overview

  

  • Entirelly Digital Design.

  • Portable and Light Weight (with handle).

  • Single Color TFT Display (Resolution: 192×64), Well Design Touch Panel.

  • Thermal Recorder (50mm×20m), Rolling Paper and Folding Paper Selectable.

  • AC/DC Power Supply. Built-in High Capacity Li-ion Battery, Support Uninterrupted Working up to 2 Hours.

  • Simultaneously 12 Leads Acquisition, Optional Selections for Rhythm Lead.

  • High-definition Digital Wave Filter to Avoid Baseline Drift, AC and EMG Interference.

  • Baseline Automatic Adjustable, Optimize Printing Position.

  • High Resolution (vertical 8dot/mm, horizontal 40dot/mm) Thermal Printer, ECG Waveforms, Lead Name, Paper Feed Speed, Gain, Analysis Report, Patient’s information can be Printed.

  • Automatic Measurement of ECG Parameters with Analysis Report, Various Printout Format.

  • 120 Seconds ECG Waveform Replay, 60 Exams Patient Data Storage.

  • Memory Expansion Capacity by Connecting with USB Flash Disk.

  • Accurate Indication for Unstable Electrode Connection, Lead-off Detection and Display.

  • Complete ECG Amplifier Grounding, Enhance the Safety of Patient.

  • Network Interface Provide Far-distance Data Transporting Capability.

  • Standard Accessories: 1 x Patient Cable, 6 x Chest Electrodes, 4 x Limb Electrodes, 1 x Paper Roller, 1 x Record Paper, 1 x Ground Line, 1 x Input/Output Signal Cable.

 

Specifications

  

Lead Standard

12 Leads

Display

Color TFT Display (Resolution: 192×64) and Touchpanel

Rhythm Lead

Optional 12 leads

Input Circuit

Isolated, Protection against pace-making and defibrillation

Operation Mode

Automatic/Manual

Recording Paper

Roll or Folded, for 1-channel 50mmX20m

Input Impedance

≥ 10MO

Frequency Response

0.05-150Hz (-3dB)

Input Circuit Current

≤ 50nA

Recording Speed

25,50mm/s

Gain

5,10,20mm/mV

Filter AC

50/60Hz—20dB

EMG

35/45Hz—3dB

Sampling Rate

1kHz,12 leads

A/D Converter

12bit

Skin Voltage Tolerance

± 350mV

CMRR

≥ 90dB

1 wave Display

192×64pixels LCD TFT Display

Analysis Function

Output analysis

Calibrating

1mV ± 2%

Calibrating

1mV ± 2%

Time Constant

3.2s (0, +20%)

Noise Level

≤ 15uVp-p

External Input

10mm/V±5%, single ended ≥100kΩ

ECG Output

1V/mV±5%, single ended ≤100Ω

Baseline Control

Automatic position

Recording System

Thermal array printer

Memory

60 exams

Safety Level

Class 1, Type CF

Wide range power supply

110/220V AC, 50/60Hz

Downloads

Brochure:

AV-9100 (Single Channel ECG)